Online Wall-Mounted Trace Oxygen Analyzer (CI-PC95)
-
Microprocessor-based core design, delivering excellent long-term stability and high reliability.
-
Uses an electrochemical sensor as the measuring unit, providing high precision and a long calibration interval.
-
High-precision automatic temperature and pressure compensation, reducing the impact of environmental variations on measurement accuracy.
-
Overpressure protection: when gas path pressure exceeds the protection range, the system automatically cuts off the gas path on both sides of the sensor chamber.
Principle: Electrochemical
Measuring range: 0~10/100/1000ppm/1/25.00% O2
Accuracy:
0.01~9.99ppm O2: ±5%FS
10.0~99.9ppm O2: ±3%FS
100~1000ppm O2: ±2%FS
Repeatability:
0.01~9.99ppm O2: ±2.5%FS
10.0~99.9ppm O2: ±1.5%FS
100~1000ppm O2: ±1%FS
Stability:
0.01~9.99ppm O2: ±2.5%FS/7d
10.0~99.9ppm O2: ±1.5%FS/7d
100~1000ppm O2: ±1%FS/7d
Response time: T90<60s (25°C)
Power Supply: DC 24V
Communication: RS485(standard)/RS232(optional)
Ambient Temperature: -10~+45°C
Ambient Humidity: <90%RH
Switching Signal: 5 groups,control signal DC 9~28V
Sample Gas Flow: 0.5~2L/min (recommended 1.5L/min)