Portable Trace Oxygen Analyzer (CI-PC931/932)
-
High precision temperature and pressure automatic compensation, reduce the impact of temperature and pressure on the measurement accuracy;
-
With overpressure protection, when the pressure in the gaspath exceeds the pressure protection range, automatically cut off the gas path on both sides of the sensor chamber;
-
The use of digital sensors, convenient for users to replace on site, device is ready to be used once the sensor is replaced.
Principle: Electrochemical
Measuring range: 0~10/100/1000ppm/1/25.00% O2
Accuracy:
0.01~9.99ppm O2: ±5%FS
10.0~99.9ppm O2: ±3%FS
100~1000ppm O2: ±2%FS
Repeatability:
0.01~9.99ppm O2: ±2.5%FS
10.0~99.9ppm O2: ±1.5%FS
100~1000ppm O2: ±1%FS
Stability:
0.01~9.99ppm O2: ±2.5%FS/7d
10.0~99.9ppm O2: ±1.5%FS/7d
100~1000ppm O2: ±1%FS/7d
Response Time: T90<60s (25°C)
Power Supply: AC 100~240V, 50/60Hz
Communication: RS485(standard)/RS232(optional)
Analog Signal: 1 group,4~20mA/0~20mA/0~1V/0~5V/0~10V/1~5V, can be set freely
Working Temperature.: 0~+45°C
Ambient Humidity:<80%RH
Sample Gas Pressure: Atmospheric Conditions±10% (Gas outlet should be atmospheric conditions)