GNL-B1A Portable Trace Oxygen Analyzer
-
With overpressure protection, when the pressure in the gas path exceeds the pressure protection range, automatically cut off the gas path on both sides of the sensor chamber;
-
High precision temperature and pressure automatic compensation, reduce the impact of temperature and pressure on the measurement accuracy;
-
Unique solenoid valve protection gas path: built-in solenoid valve protection gas path, the protection valve can reopen unlimited times and have a long service life;
-
Microcontroller controls the analyzer to be in bypass or sample mode.
Principle: Electrochemical
Measuring range: 0~10/100/1000ppm/1/25.00% O2
Accuracy:
0.01~9.99ppm O2: ±5%FS
10.0~99.9ppm O2: ±3%FS
100~1000ppm O2: ±2%FS
Repeatability:
0.01~9.99ppm O2: ±2.5%FS
10.0~99.9ppm O2: ±1.5%FS
100~1000ppm O2: ±1%FS
Stability:
0.01~9.99ppm O2: ±2.5%FS/7d
10.0~99.9ppm O2: ±1.5%FS/7d
100~1000ppm O2: ±1%FS/7d
Response Time: T90<60s (25°C)
Recovery Time: It takes 60 minutes to go from air (20.94%) to 10 ppm
Measuring Medium: Nitrogen, Hydrogen, Helium, Argon, and Hydrocarbon gases
Power Supply:
When equipped with a transformer: 220VAC ±10%, 50/60Hz
When the switching power supply is configured: 85~264VAC, 50/60Hz