CI-PC121 Online Oxygen Analyzer
Gas path pressure detection function: real-time detection analyzer gas path pressure, easy to judge whether the gas pathblocked;
Window surface covered with a toughened glass, enhance the impact protection of the window;
Capacitive touch key, key sensitive, reliable, simple oрeration;
Three-way alarm switch output, compact size, simple structure, easy to carry and maintain.
Principle: Frequency Conversion Ion Flow
Measuring range: 0~25.00% O2
Accuracy:
0.01~99.9ppm O2: ±10ppm
100~999.9ppm O2: ±3%FS
1000~9999ppm O2: ±3%FS
1.00%~25.00% O2: ±3% of measured value
Repeatability:
0.01~99.9ppm O2: ±5ppm
100~999.9ppm O2: ±1.5%FS
1000~9999ppm O2: ±1.5%FS
1.00%~25.00% O2: ±1.5% of measured value
Stability:
0.01~99.9ppm O2: ±5ppm/7d
100~999.9ppm O2: ±1.5%FS/7d
1000~9999ppm O2: ±1.5%FS/7d
1.00%~25.00% O2: ±1.5% of measured value/7d
Response Time: T90<60s (After full preheating)
Power Supply: AC 100~240V, 50/60Hz
Communication: RS232(standard)/RS485(optional)
Ambient Temperature: -10~+50°C;
Ambient Humidity: <80%RH