CI-PC125 Online Oxygen Analyzer
Electronic flowmeter, accurate flow detection, long service life;
The analyzer has passed strict anti-electromagnetic interference and static tests before leaving the factory, reaching the military standards;
When the oxygen concentration deviates from the alarm value, the alarm value can be set in the full range of arbitrary.
Principle: Frequency Conversion Ion Flow
Measuring range: 0~25.00% O2
Accuracy:
0.01~99.9ppm O2: ±10ppm
100~999.9ppm O2: ±3%FS
1000~9999ppm O2: ±3%FS
1.00%~25.00% O2: ±3% of measured value
Repeatability:
0.01~99.9ppm O2: ±5ppm
100~999.9ppm O2: ±1.5%FS
1000~9999ppm O2: ±1.5%FS
1.00%~25.00% O2: ±1.5% of measured value
Stability:
0.01~99.9ppm O2: ±5ppm/7d
100~999.9ppm O2: ±1.5%FS/7d
1000~9999ppm O2: ±1.5%FS/7d
1.00%~25.00% O2: ±1.5% of measured value/7d
Response Time: T90<60s (After full preheating)
Power Supply: AC 100~240V, 50/60Hz
Communication: RS232(standard)/RS485(optional)
Ambient Temperature: -10~+50°C;
Ambient Humidity: <80%RH
Sample Gas Flow: 0.5~2.0L/min (Recommended 1.5L/min)