Trace Moisture Transmitter (CI-PC382)
The combination of microcontroller and digital processing technology, anti-cross interference, stable and reliable;
Using ultra-thin alumina capacitive sensor as measuring unit, high precision, long calibration cycle;
RS485 communication mode: Isolated analog signal output, support software switching 0-20mA/4-20mA analog output mode.
Principler: Electrolysis
Measuring Range: 0~500ppmV H2O (can be customized according to customer requirements)
Linear Error: 0~10ppm ±3%FS, 10~500ppm ±5%FS
Repeatability: 0~10ppm ±1.5%FS, 10~500ppm ±2.5%FS
Stability: 0~10ppm ±1.5%FS/7d, 10~500ppm ±2.5%FS/7d
Response Time: T63<3min
Coating Period: 6 months, depending on use, the higher the concentration measured, the shorter the coating period
Calibration Period:2 years (recommended)
Power Supply: DC 24V
Transmitter Power: <15VA
Ambient Temperature: -10~+50°C
Ambient Humidity: <90%RH
Flow Rate: 20L/h(there must be flow rate when measuring)