Online Selective Hydrogen Analyzer (CI-PC502)
The sensor is specific to hydrogen and has no cross-sensitivity to other gases;
The surface of the window is covered with a piece of toughened glass to enhance the impact protection of the window;
Using capacitive touch buttons, the buttons are sensitive, reliable and easy to operate;
Menu Lock Switch: prevent misoperation from changing analyzer parameters and affecting its normal operation;
Convenient data storage function, support U disk and SD card, large data storage capacity.
Measuring Range: 0.01%~10%, 15%~99.99% H2 (customizable)
Accuracy: 0.3%
Repeatability: 0.2%
Stability: 0.2%/7d
Linearity: 0.2%
Background Gas: N2
Response Time: 30s
Power Supply: AC 85~264V, 50/60Hz
Consumption: <25VA
Ambient Temperature: Gas: -20~+60°C
Operation: -10~+55°C
Storage: -20~+60°C
Ambient Humidity:<90%RH
Sample Gas Pressure: Atmospheric Conditions ±10% (Gas outlet should be atmospheric conditions)
Sample Gas Flow: 0.1~10slpm (1.0slpm recommended)
Pump: Required for negative pressure; slight positive pressure optional.