CI-PC952-1 Online Hydrogen Sulfide Gas Analyzer
Microprocessor-based design ensures excellent long-term stability and high reliability;
Equipped with high-precision electrochemical gas sensor with extended calibration intervals;
Supports data storage via USB flash drive and SD card, offering large capacity for easy data processing and analysis;
Built-in timed data logging allows quick and convenient access to historical test data;
Principle: Electrochemical
Measuring range: 0~10000ppm H2S
Resolution: 0.1ppm
Accuracy: ±2%FS
Repeatability: ±1%FS
Stability: ±1%FS/7d
Response Time: T90<35s (0~20ppm)
Recovery Time: <60s
Measurement Method: Online
Calibration Cycle: 1 year (recommended)
Weight: Net weight 5.0kg, packing approx.15kg
Power Supply: DC 24V
Consumption: <15VA
Communication: RS485 (standard)/RS232 (optional)
Alarm Output: Alarm point: 1
Relay Contact Capacity: DC 24V, 0.2A
Analog Signal: 1 Group, 4~20mA/0~20mA/0~1V/0~5V/0~10V/
1~5V, can be set mode freely