CI-PC562 Online Trace Hydrogen Analyzer
When the hydrogen concentration is higher than the protection concentration, the air pump is automatically turned on to feed air into the sensor; the sensor must try to avoid working in an environment with a hydrogen concentration higher than 1% for a long time to prolong its service life;
When the hydrogen concentration deviates from the alarm value, an alarm will be triggered, in which the value can be set arbitrarily within the full scale range.
Principle: Electrochemical
Measuring Range: 0~10ppm/2000ppm/2.000% H2
Linear Error: ±2%FS
Repeatability: ±1%FS
Stability: ±1%FS/7d
Response Time: T90<70s (0~1000ppm)
Power Supply: AC 100~240V, 50/60Hz
Ambient Temperature: -10~+50°C
Ambient Humidity: <80%RH
Sample gas pressure: Atmospheric conditions ±10% (Gas outlet should be atmospheric conditions)
Sample gas flow: 400~600ml/min
Alarm output: 2 groups (COM and NO)
Communication: RS485 (standard)/RS232 (optional)
Analog Signal: 1 group, 4~20mA/0~20mA/0~1V/0~5V/0~10V/
1~5V, can be set mode freely
Sensor Life: More than 2 years (normal use)
Instrument Life: More than 5 years (normal use)