Online Infrared Gas Analyzer (CI-PC261)
Online Infrared Gas Analyzer (CI-PC261)
-
Using NDIR (non-dispersive infrared measurement), the sensor is not in contact with the measured gas, and the poisoning of the sensor will not occur;
-
It has the characteristics of intelligence, good stability, high reliability, and long calibration cycle;
-
Menu Lock Function: Prevent misoperation from changing analyzer parameters and affecting its normal operation;
-
Two-way single-pole single-throw alarm function, the alarm value can be set freely within the full range.
Principle: Non-dispersive Infrared (NDIR)
Measuring Range:
| Components | Min. Range | Max. Range |
| CO | 0~2% | 0~100% |
| CO2 | 0~2000 ppm | 0~100% |
| CH4 | 0~4.4% | 0~100% |
| SF6 | 0~50 ppm | 0~1000 ppm |
Accuracy: ±2% FS (±1%FS optional)
Repeatability: ±1% FS
Zero drift: ±1% FS/7d
Range drift: ±1% FS/7d
Response time: T90<15s (500ml/min)
Power supply: AC 100~240V, 50/60Hz
Ambient Temperature: 5~+45°C
Ambient Humidity: <90%RH
Sample Gas Pressure: Atmospheric Conditions ±10% (Gas outlet should be atmospheric conditions)
Communication: RS485 (standard)/RS232 (optional)
Analog Signal: 1 group, 4~20mA/0~20mA/0~10V, can be set mode freely